ACR Electronics SmartReader Plus Specifications Page 58

  • Download
  • Add to my manuals
  • Print
  • Page
    / 117
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 57
SmartReader Plus Data Loggers
49
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.9.5 Manual Calibration
A simple way to manually adjust the calibration values on a SmartReader Plus 6 thermocouple
logger follows. For each active channel:
1. Install the thermocouple and set the proper equation number for that thermocouple
type. Set the TrendReader software into Realtime mode.
2. Using a valid low reference temperature (as low as possible), observe the Realtime
reading and adjust the Low calibration value until the logger displays the reference
temperature.
3. Using a valid high reference temperature (as high as possible), observe the Realtime
reading and adjust the Mid calibration value until the logger displays the reference
temperature.
To obtain the best results from the calibration, set the reference temperatures as far apart as
possible.
Page view 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 116 117

Comments to this Manuals

No comments